very obvious in 666986_2023-08-16 block 5:
666986_2023-08-16
block 5: trials 368 - 458
block 6: trials 459 - 552
long frame interval (0.04s) in task hdf5 trial 421, frame 165707 (60% through block 5)
affects vis and aud stimuli
affects all probes recorded
sync processing otherwise appears normal, no warnings
barcode intervals appear normal on sync and ephys
does not affect lick times
check drift map (currently broken figurl components broken AllenNeuralDynamics/aind-ephys-pipeline#20)
The text was updated successfully, but these errors were encountered:
This is a probe-sync issue (~200 ms of data dropped in ephys recording):
obs_intervals
Sorry, something went wrong.
No branches or pull requests
very obvious in
666986_2023-08-16
block 5:block 5: trials 368 - 458
block 6: trials 459 - 552
long frame interval (0.04s) in task hdf5 trial 421, frame 165707 (60% through block 5)
affects vis and aud stimuli
affects all probes recorded
sync processing otherwise appears normal, no warnings
barcode intervals appear normal on sync and ephys
does not affect lick times
check drift map (currently broken figurl components broken AllenNeuralDynamics/aind-ephys-pipeline#20)
The text was updated successfully, but these errors were encountered: