Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

Timing issue evident in PSTH #121

Open
1 task
bjhardcastle opened this issue Sep 17, 2024 · 1 comment
Open
1 task

Timing issue evident in PSTH #121

bjhardcastle opened this issue Sep 17, 2024 · 1 comment

Comments

@bjhardcastle
Copy link
Member

very obvious in 666986_2023-08-16 block 5:
fig3c_666986_2023-08-16_F-213

  • block 5: trials 368 - 458

  • block 6: trials 459 - 552

  • long frame interval (0.04s) in task hdf5 trial 421, frame 165707 (60% through block 5)

    • not exactly aligned with shift
  • affects vis and aud stimuli

    • not a stim alignment issue (same issue when plot is aligned on trial start)
  • affects all probes recorded

  • sync processing otherwise appears normal, no warnings

  • barcode intervals appear normal on sync and ephys

  • does not affect lick times

  • check drift map (currently broken figurl components broken AllenNeuralDynamics/aind-ephys-pipeline#20)

@bjhardcastle bjhardcastle added this to the paper1-datacube milestone Sep 26, 2024
@bjhardcastle
Copy link
Member Author

bjhardcastle commented Oct 4, 2024

This is a probe-sync issue (~200 ms of data dropped in ephys recording):
666986_2023-08-16

  • method for getting sampling rate needs updating
  • dropped periods need to be cut, and marked with obs_intervals in nwb